Medical device risk management specialist implementing ISO 14971 throughout product lifecycle. Provides risk analysis, risk evaluation, risk control, and post-production information analysis. Use when user mentions risk management, ISO 14971, risk analysis, FMEA, fault tree analysis, hazard…
Why we recommend it: ISO 14971 medical device risk management, adjacent to healthcare quality and safety.
Inspect before use
Read the source instructions and check the requirements before you install this skill.
SKILL.md
The saved excerpt is a snapshot from when this skill was listed. The external source remains the complete and most current version.
---
name: "risk-management-specialist"
description: Medical device risk management specialist implementing ISO 14971 throughout product lifecycle. Provides risk analysis, risk evaluation, risk control, and post-production information analysis. Use when user mentions risk management, ISO 14971, risk analysis, FMEA, fault tree analysis, hazard identification, risk control, risk matrix, benefit-risk analysis, residual risk, risk acceptability, or post-market risk.
---
# Risk Management Specialist
ISO 14971:2019 risk management implementation throughout the medical device lifecycle.
---
## Table of Contents
- [Risk Management Planning Workflow](#risk-management-planning-workflow)
- [Risk Analysis Workflow](#risk-analysis-workflow)
- [Risk Evaluation Workflow](#risk-evaluation-workflow)
- [Risk Control Workflow](#risk-control-workflow)
- [Post-Production Risk Management](#post-production-risk-management)
- [Risk Assessment Templates](#risk-assessment-templates)
- [Decision Frameworks](#decision-frameworks)
- [Tools and References](#tools-and-references)
---
## Risk Management Planning Workflow
Establish risk management process per ISO 14971.
### Workflow: Create Risk Management Plan
1. Define scope of risk management activities:
- Medical device identification
- Lifecycle stages covered
- Applicable standards and regulations
2. Establish risk acceptability criteria:
- Define probability categories (P1-P5)
- Define severity categories (S1-S5)
- Create risk matrix with acceptance thresholds
3. Assign responsibilities:
- Risk management lead
- Subject matter experts
- Approval authorities
4. Define verification activities:
- Methods for control verification
- Acceptance criteria
5. Plan production and post-production activities:
- Information sources
- Review triggers
- Update procedures
6. Obtain plan approval
7. Establish risk management file
8. **Validation:** Plan approved; acceptability criteria defined; responsibilities assigned; file established
### Risk Management Plan Content
| Section | Content | Evidence |
|---------|---------|----------|
| Scope | Device and lifecycle coverage | Scope statement |
| Criteria | Risk acceptability matrix | Risk matrix document |
| Responsibilities | Roles and authorities | RACI chart |
| Verification | Methods and acceptance | Verification plan |
| Production/Post-Production | Monitoring activities | Surveillance plan |
### Risk Acceptability Matrix (5x5)
| Probability \ Severity | Negligible | Minor | Serious | Critical | Catastrophic |
|------------------------|------------|-------|---------|----------|--------------|
| **Frequent (P5)** | Medium | High | High | Unacceptable | Unacceptable |
| **Probable (P4)** | Medium | Medium | High | High | Unacceptable |
| **Occasional (P3)** | Low | Medium | Medium | High | High |
| **Remote (P2)** | Low | Low | Medium | Medium | High |
| **Improbable (P1)** | Low | Low | Low | Medium | Medium |
### Risk Level Actions
| Level | Acceptable | Action Required |
|-------|------------|-----------------|
| Low | Yes | Document and accept; still reduce as far as possible (EU MDR) |
| Medium | After reduction AFAP | Reduce as far as possible; document why further reduction is impossible |
| High | After reduction AFAP | Reduction required; demonstrate all further options exhausted |
| Unacceptable | No | Design change mandatory |
> **EU MDR — AFAP, not ALARP:** For CE-marked devices, risks must be reduced **as far as possible (AFAP)** without economic considerations (MDR Annex I, GSPR 1–4; EN ISO 14971:2019/A11:2021 Z-annexes deviation). ALARP ("as low as reasonably practicable"), which permits cost-benefit weighing in acceptability decisions, is **not an acceptable criterion under the EU MDR** — a notified body will flag it. ISO 14971:2019 itself removed ALARP from the normative text. ALARP may persist in some non-EU jurisdictions (e.g., the UK HSE tradition); if used outside the EU, flag the deviatiRead the full source on GitHub (opens external page)